Menu

Water-Trap with extra fine particle filter

model WT 20.5
 

- Secure protection of the gas analyser from condensate and extra fine dust
- Low costs for reliable safety

 

Functional description:

The Water-Trap is installed directly in front of the gas analyser in the piping. If the upstream gas processing (cooler, perestaltic pump, dust filter and the like) fails, the Water-Trap uses its semi-permeable membrane with SUN-C coating to protect the gas analyser.
The diaphragm separates gases from condensate and extra fine dust. If it is fully filled, the gas flow will be interrupted. An alarm is indicated via the upstream flowmeter with monitoring (refer to sample applications).

The Water-Trap is also known as a “Police filter”

Additional information from experience:

If the filter element, of the dust filter located in the gas processing is changed, particles are consistently dissolved and directed into the analyser. The downstream Water-Trap prevents this contamination as well with its diaphragm pore size of 0.1 μm.

Technical specifications:

Scope of delivery: Water-Trap, connection adapters (option), assembly bracket (option)
Water pressure membrane: 0 - 2 bar
Operating pressure for gas: 0 - 2 bar
Gas flow: 0 - 400 l air/h
Pressure drop at 100 l air/h: approx. 10 mbar
Pressure drop at 400 l air/h: approx. 40 mbar
Diaphragm pore size: < 0,1μm
Operating temperature: 0°C to +90°C
Effective filter area: 25 cm2
Housing volume: 5 ml
Materials used: PTFE, PP, SUN-C coating
Dimensions: Diameter 70 mm, length 120 mm
Gas connections: on both sides 1/8“ NPT outside thread or on both sides 6 mm pipe nozzle or on both sides 6-12 mm stepped hose barb
Assembly: Mounting in the existing piping
Operating instructions in German and English (included in the scope of delivery)
  Spanish, Italian, French, Russian upon request



Suitable for the following applications:

  • Emission measurements with fossil fuels
  • Waste incineration plants
  • Biogases
  • Ambient air monitoring
  • Silo monitoring
  • Heat treatments
  • Cement, glass, steel, paper industries
  • Combustion engines

Not suitable for the following applications:

Condensate with aromatic hydrocarbons, e.g. oils and fuels in refinery process gases. The models 20.82 E and the model 20.83 E XL with their oil-block membrane and coalescence filters have been designed for this.

Protection from dust:

The Water-Trap shows severe contamination of dust. The diaphragm with its pore size of 0.1 μm retains even the finest dusts.

Connection techniques:

Order numbers:

Item Item number
Water-Trap WT 20.5 R, connections on both sides 6 mm nozzle WT205R
1 pair (2 units) connection adapter with 6 mm pipe connection for screwed hose connection 6/4 mm VE6R64
Water-Trap WT 20.5 N, connections on both sides 1/8” NPT outside thread WT205N
1 pair (2 units) connection adapter with 1/8” NPT threaded connection for screwed hose connection 6/4 mm VE18N64
Water-Trap WT 20.5 S, connections on both sides 6.1 mm-12.7 mm stepped hose barb WT205S
Assembly brackets for wall mountig for all models of the WT 20.5 MONWIWAMO205

 

Sample application 1: Flue gas analysis

Advantage:

The analyser is protected in case of failure of the cooling system (cooler, perestaltic pump) Another Water-Trap can be used for protecting the gas processing.

Sample application 2: TOC measurement

Advantage:

The analyser is protected in case of failure of the cooling system (cooler, perestaltc pump).

Sample application 3: Strip systems

Advantage:

In case of a clogged flow, the HC-FID is protected from harmful water.